Osmometry & applications

Freezing point or vapour pressure osmometer for your sample?

Start with the required method and the sample itself. Compare OSMOMAT and VAPRO against the solvent, volatile components, viscosity and available volume.

On this page
Gonotec Osmomat 3000D
Gonotec Osmomat 3000D

Which method will the customer accept for reporting osmolality? If their SOP specifies freezing point depression, choose an instrument that supports it. Where the method is open, note the solvent, volatile components such as alcohol, viscosity and available volume before arranging a trial.

Osmolality is expressed per mass of solvent, usually in mOsm/kg. Osmolarity is expressed per volume of solution, for example mOsm/L. Check the units on the comparison report before interpreting a difference. [1, 4]

Which samples suit an OSMOMAT freezing point osmometer?

After calibration, the Gonotec OSMOMAT 3000 determines the osmolality of aqueous solutions from their freezing point depression. Its manual specifically excludes organic, saturated and highly viscous solutions. Match the sample and expected measurement range to the model offered. [1]

Which samples suit a VAPRO vapour pressure osmometer?

The VAPRO 5600 measures dew-point temperature depression without freezing the sample. The manual discusses viscous samples and tissue. Sample placement and reaching equilibrium matter for those applications. For limited samples, allow for holder volume, repeats and transfer losses. [2]

Pay particular attention to ethanol or methanol. Walker and colleagues found a low response with vapour pressure osmometry over the alcohol concentrations they tested. The VAPRO manual also discusses the response to ethanol. In water–ethanol mixtures, the freezing point response can itself be disproportionate to the number of ethanol particles. If alcohol matters to your decision, compare results for a representative composition and range. An osmolality reading alone cannot identify the solute. [2, 3]

For a discussion with AE, describe the solvent, volatile components, sample condition, expected range and volume available for all repeats. Include the comparison method and any qualification requirements. The full formulation can stay confidential.

Sources and further reading

  1. Gonotec · OSMOMAT 3000basic / 3000 user guide, v1.23 Rev.1, 24 April 2019, sections 2.2 and 3.2 (pp. 14, 27). Product page
  2. ELITechGroup · VAPRO 5600/5600XR manual 57-0006-03D, revised 26 October 2022, section 1.3 and Appendix F (pp. 105–108). Documentation centre
  3. Walker et al. · The missing gap, Archives of Internal Medicine 146(9), 1986
  4. IUPAC Gold Book · Osmotic concentration, online edition 2025

Share this article

Choose equipment by application

Osmometry

Compare OSMOMAT and VAPRO for freezing-point or vapor-pressure measurement. Check sample suitability, consumables and AE support in Indonesia.

Open application guidance